Medical Power Adapter 6W UK
Features
DIMENSION (mm)
| MODEL No. | AC PIN | L | W | H |
| GS2B-006-zzz-A | UK | 64.0 | 46.0 | 48.6 |
SPECIFICATIONS
| Input | Rated Input Voltage | 100VAC-240VAC | |
| Rated Input Frequency | 50/60Hz | ||
| Input Current | 0.2A MAX. | ||
| Inrush Current | 30A MAX. | ||
| Leakage Current | 0.1mA MAX. | ||
| Standby Power | 0.3W MAX. | ||
| Output | Output Voltage | DC5-15V , DC24V | |
| Line Regulation | ±2% | ||
| Load Regulation | ±5% | ||
| Output Current | 0.01A-1.2A | ||
| Maximal Power | 6W Max | ||
| Protection | Over-Current Protection ; Short-Circuit Protection ; Over-Voltage Protection | ||
| Standard | EMC | CE: EN 60601-1-2 CISPR 11 IEC61000-3-2 IEC61000-3-3 IEC 61000-4-2,3,4,5,6,8,11 FCC: 47 CFR FCC Part 15 Subpart B ANSI C63.4 |
|
| Safety | CB:IEC 60601-1 TUV: IEC/EN 60601-1 IEC/EN 60601-1-6 IEC/EN 62366-1 UL: AAMI ES 60601-1 CSA C22.2 No.60601-1 |
||
| Environmental | Operation Temperature | 0℃ To 40℃ Ambient | |
| Storage Temperature | -20℃ TO 60℃ | ||
| High-Reliability: MTBF > 100000 Hours Under 25℃ | |||